Percussionaire Corporation,News Articles
Advanced Cardiopulmonary Dynamics
Dr. Forrest M. Bird President Bush Award President Obama Award
Main About Us Products Contacts Distributors Educational Pages News/Articles Seminars
Additional Information for:
Patients
Physicians
Insurance Providers
 
HC Impulsator

HC Impulsator® is available to Limited Release only in the U.S.
Click to view the training video

Home Care Services
(Patient Support)

Valarie Zimmerman:
208- 263-2549 Ext 161
Email: homecare@percussionaire.com

NEW! AARC Discussion group on VDR
Click to View>>
 
News Breaks
Nightline Video shows IPV-1C® being used to treat H1N1 patients
In the first section you can see an IPV-1C® by the patient and in the second the patient is being given a treatment.
Click here to see the segment>>

Concerns continue over H1N1
To hear Respiratory Care & Sleep Medicine's H1N1 Flu Roundtable
Click here >>

H1N1: IPV improves on HFOV in Central Illinois Hospital< <click here > >


World Expert on Ventilators Collaborates With Upstate Researchers!!
click here to view the whole story!

IPV Related Studies
To view articles and studies involving IPV
click here

Bookmark and Share

Percussionaire® Corporation Educational Documents

A COMPENDIA OF THE MECHANICAL VENTILATION OF THE LUNG AS IMPACTED UPON, BY THE LIFE LONG CONTRIBUTIONS OF FORREST M. BIRD, M.D., PhD., ScD.

Latest News as @ Monday, September 06, 2010
Educational Documents » Part 8 »
HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV™) -

Note: For maximal comparative understanding, FIRST read PART SEVEN which defines the concept of High Frequency Oscillatory Ventilation (HFOV).

The fluidic logic employed in the design of devices capable of High Frequency Percussive Oscillatory Ventilation (HFPOV™), allows a major reduction in transition penalties when compared to HFOV devices. The primary device for HFPOV™ is called an Oscillatron®, which has cyclic transition penalties at 900 cycles per minute of about 14 milliseconds compared to piston and magnetically servoed HFOV ventilators with transition penalties approaching 42 milliseconds. By subtracting 14 from 42 one can see that the time for effective reciprocating tidal exchange gradients is increased by 28 milliseconds in the Oscillatron® fluidic HFPOV™ design.



It follows that a reduction in the high frequency delivery frequencies will result in more time for tidal exchange by any dynamic (push-pull) oscillator design. However, without the intrapulmonary gas-mixing enhancement provided by percussive endobronchial gas mixing, the potential for CO2 recruitment will be minimal.

By modulating the tidal delivery waves of certain CMV ventilators, the clinical efficacy can be increased with a potential reduction in PIP and FIO



It follows that increasing the blood/gas interface by mechanically increasing the D/FRC with an elevated CPAP alone could serve to increase PaO2. However, without a concomitant tidal exchange sufficient to "wash out" CO2, the patient could die (nice and pink) from a CO2 narcosis.

THE U S FDA CLEARED THE PERCUSSIONAIRE® OSCILLATRON® FOR MARKETING ON OCTOBER 16, 1989.




OVER THE MANY YEARS, THE VOLUMETRIC DIFFUSIVE RESPIRATION CONCEPT (VDR®) HAS TAKEN PRECEDENCE OVER THE OSCILLATRON® AS A UNIVERSAL VENTILATOR.

On October 26, 1990, the FDA provided Dr. Bird with the only 510K frequency increase for high frequency oscillatory ventilators.



This allowed other piston and magnetically servoed high frequency oscillators to increase their cyclic frequency, based upon Dr. Bird’s prior state of the art available on May 28, 1976.

Essentially, the "Oscillatron 1" defines the extension of the (positive phase) Phasitron® enabling a potential sub ambient expiratory phase.

The clinical concept of Volumetric Diffusive Respiration (VDR®) has confirmed the clinical efficacy of combining "Diffusive Percussive Ventilation" with the traditional Convective CMV ventilation of the lung. Therefore, it became a natural timely transition to enable standard conventional CMV ventilators to increase their level of clinical efficacy by combining "Diffusive Intrapulmonary Percussive Ventilation" with the CMV programmed convective tidal exchange. The Oscillatron® alone (among all other HFOV ventilators) enables the modulation of a conventional tidal exchange (with PIP’s approaching 100 cm H2O), providing for the enhanced diffusion of the intrapulmonary gases in lungs with very low compliance.

The Oscillatron® is essentially a unique Canister which is servoed by the Phasitron to generate a potential sub ambient phase during High Frequency Percussive Oscillatory Ventilation (HFPOV™).



The fluidic design employed in the percussive Oscillaton® logic is advanced by a "Differential Flow Interrupter Cartridge" with near instantaneous opening and closing thus providing repetitive flow interruption to an encapsulated sliding venturi injector exhalation valve called a Phasitron®".

In turn, the Phasitron® near instantaneously pressurizes an encapsulated diaphragm centered in the chamber by opposing counter springs. The percussive forward movement of the diaphragm injects a high velocity tidal volume of gas into the breathing circuit to modulate a selected positive pressure at the patient’s proximal airway. The inspiratory percussive flow gradient is near instantly reversed, creating an immediate follow on expiratory flow gradient, which generates a potential distal–proximal sub ambient pressure gradient at the proximal airway.





The advantage of employing a fluidic "flow interruption" logic as opposed to "piston and magnetic flow generations" includes a number of factors, encompassing:

1.  A virtually unlimited source of regulated high pressure gas exerts continuous pressures of about 40 psig against the flow interrupter valve gate. When the flow interrupter cartridge valve gate near instantaneously opens, a sustained percussive burst of regulated high pressure gas is released against the jet of the Phasitron® venturi for a timed interval.

2.  In turn, the velocity of the flow leaving the distal end of the venturi tube is amplified and accelerated.

3.  Finally, the percussive inspiratory flow generated within the venturi tube impacts upon the servoing side of the diaphragm encapsulated within the canister servoing chamber, causing a near immediate forward diaphragmatic implosion into the percussion chamber.



4.  The gas displaced within the percussion chamber is injected into the breathing circuit, creating a near instantaneous percussive pressure rise at the proximal airway. This produces an inspiratory flow gradient into the pulmonary airway.

    5.  Within about 4 milliseconds following the injection of the percussive tidal volume into the physiological airways, the inspiratory servoing pressure against the diaphragm "collapses" allowing the counter spring (previously compressed during the forward movement of the diaphragm) to expand against the diaphragm, creating a rapid volume expansion within the percussion chamber. As the diaphragm rapidly retreats out of the percussion chamber, a potential sub ambient pressure is created, establishing an expiratory flow gradient at the proximal end of the endotracheal tube.

    6.  It is the combination of a flow interruption valve releasing a gas, already under a selected compression, to the jet of a non-gated ambient referencing venturi which percussively moves the diaphragm forward, compressing the gas within the compression chamber of the canister, followed by the near instantaneous end Inspiratory pressure collapse within the venturi, which limits the inspiratory to expiratory flow gradient reversal time to about 7 milliseconds.

    The functional schematic of the Oscillatron® low inertia fluidic (push-pull) Percussive Oscillator.



    7.  At end inspiration, by the near instantaneous release of any countering venturi servoing pressure against the servoing side of the diaphragm, the compressed counter spring within the compression chamber freely recoils, moving the diaphragm backwards (out of the compression chamber) to create a near instantaneous potentially sub ambient expiratory flow gradient at the proximal end of the Endotracheal Tube.



      A MAJOR CONSIDERATION IS:

      In practice Magnetic HFOV oscillators require CPAP values up to 20 cm H2O to mechanically increase the FRC increasing the blood gas interface. This reduces the expiratory distal/proximal flow gradient, requiring a greater expiratory time period of potential sub ambient pressure to prevent a mechanical inadvertent PEEP. The main advantage of a magnetically controlled diaphragm compared to a symmetric piston oscillator, is the ability to have a positive expiratory i/e ratio such as 1:1.5 or more to provide for a greater physiological expiratory flow time. When the mandated design Transition Penalty (flow reversal penalties) time (associated with HFOV) is compounded by the positive i/e ratio, the available time for an oscillatory sub tidal exchange becomes time limited.



      When the design inertia of a Magnetic HFOV is compared with the Oscillatron’s percussive pneumatically servoed diaphragm, the transition penalty is about 14 ms. for HFPOV compared to about 42 ms. for a magnetically servoed HFOV oscillator diaphragm. The magnetic HFOV design, mandates an elevated TRANSITION PENALTY when compared to the Oscillatron’s HFPOV design. Functionally, the Oscillatron (HFPOV) with transition penalties of 14 milliseconds as opposed to 42 milliseconds for the magnetic HFOV, possesses a more effective intrapulmonary "gas exchange time" for a sub tidal gas exchange.

      Most important, the percussive Oscillatron® HFPOV design clinically requires less than half of the required scheduled CPAP of the magnetically HFOV design. High Frequency Percussive Oscillatory Ventilation (HFPOV) compared to High Frequency Oscillatory Ventilation (HFOV) has a greater sub ambient expiratory flow gradient time. Thus, with a more effective expiratory distal to proximal flow gradient a neutral i/e ratio of 1:1 is allowed without mechanically creating an inadvertent PEEP.

      8.  Uniquely, the fluidic Oscillatron® design enables successive near constant sub tidal (stroke) volume deliveries for the modulation of convective tidal intrapulmonary CMV Pressures from 5 to 100 cm H2O.

      9.  The fluidic design allows the end expiratory position of the diaphragm to compensatorily over-travel into the servoing chamber in proportion to the end inspiratory proximal airway pressure increase.

      10.  This diaphragmatic re-set allows the Oscillatron® to modulate the tidal waves of CMV ventilators with near constant sub tidal volumes, with peak delivery pressures approaching 100 cm H2O.

      11.  In order to have a clinically effective percussive injection of a tidal volume into the breathing circuit, the available injection time at each frequency must maintain a near constant flow gradient, with sufficient time to deliver a scheduled tidal exchange. Without a near immediate effective expiratory potentially sub ambient flow gradient, an inadvertent PEEP would increase with frequency, mandating an increased D/FRC.



      Another major advantage of the fluidic Oscillatron® design is that of being able to inject (near constant) high frequency tidal volumes with percussive velocities into a breathing circuit with convective tidal waves rising from near ambient to PIP’s of up to 100 cm H2O.

      When High Frequency Percussive Ventilation (HFPOV) is compared to High Frequency Oscillatory Ventilation (HFOV), the importance of the "Oscillatory Transition Penalty" becomes evident.

      The principal clinical efficacies of High Frequency Oscillatory Ventilation (HFOV) are based upon key primary factors, which are:

      1.  Increasing the blood/gas interface by means of a Continuous Positive Airway Pressure (CPAP) to increase the Dynamic Functional Residual Volume (D/FRC).

      2.  Increasing the Final Inspiratory Oxygen Concentration (FIO

        3.  Increasing the endobronchial mechanical mixing to enhance intrapulmonary diffusion by continuously delivering high frequency intrapulmonary tidal exchanges.



        It follows that:

        The less clinically effective the intrapulmonary mechanical mixing to enhance endobronchial diffusion, the greater the mandate to increase the blood/gas interface with:



        An increasing static Continuous Positive Airway Pressure (CPAP).

        An Increase in the Final Inspiratory Oxygen Concentration (FIO2).

        Therefore, the greater the high frequency pulsed tidal exchange volume, the less the required CPAP and FIO2 to maintain acceptable PaO2 and PaCO2.

        The higher the mandated CPAP, the higher the mean intrathoracic pressure.

        The greater the increase in the mean intrathoracic pressure, the greater the induced pulmonary arterial pressure.

        The greater the pulmonary arterial pressure, the greater the right heart strain and/or failure.

        Therefore, the pulsatile (oscillatory) exchange limits of the mechanical design of the HFOV ventilator will determine the clinical efficacy of the HFOV concept

        Technologically, it is the high transition penalty mandated by the design of piston and magnetic HFOV oscillators that serves to inhibit their clinical efficacy. The fluidic dynamic (push-pull) percussive Oscillatron® design reduces the Oscillation transition penalty from about 42 milliseconds for HFOV to about 14 milliseconds for HFPOV™.

        This major reduction in the HFPOV™ transition penalties allows a more effective intrapulmonary gas mixing, greatly reducing the required CPAP and FIO2 requirements of HFPOV™, and effectively reducing the mean intrathoracic pressure and the imposition upon right heart function". HFOV requires CPAP values approaching 20 cm H2O.

        HFPOV requires less than half of these CPAP values. Therefore, HFOV requires a positive i/e ratio while HFPOV with CPAP values under 10 cm H2O, requires only a neutral 1:1 i/e ratio with the greater Proximal/Distal pulmonary airway flow gradients.



        IN SUMMARY, THE OSCILLATRON® CAN PROVIDE CERTAIN CMV VENTILATORS WITH INCREASED CLINICAL EFFICACY BY INCREASING THEIR ENDOBRONCHIAL MECHANICAL DIFFUSION.





        THE OSCILLATRON® SERVO FOR INTERFACING CMV VENTILATORS



        This educational program was prepared and edited by

        © Bird Institute of Biomedical Technology, 2005

        Related Documents:
        Download Printable PDF http://www.percussionaire.com/pdfs/part8.pdf

        « Go Back



        General Information

        General IPV® Manual - click here >> (right click and save as to download)

        Percussionaire® Product Technical Data - click here >> (right click and save as to download)

        Bronchitis General Manual Of Understanding - click here >> (right click and save as to download)

        The History Of Evidence Based Medicine - click here >> (right click and save as to download)

        The History Of Percussionaire® - click here >> (right click and save as to download)

         