Battery operated DISCONNECT MONITORING, is located on right facing side of VDR-4 housing. Toggle Switch to ON for Starting, with Onset Time control knob Arrow under the 12:00 index.
Sustained pressure rise alarming is determined by Red SENSITIVITY knob selection, normally with the control knob Arrow at the 12:00 position a sustained pressure of over 100 cm H2O has to be maintained for over two (2) seconds to activate alarming and associated proximal airway pressure drop.
FOR DIGITALIZED VENTILATOR MONITORING- PROGRAM MONITRON® II.
Monitron- ON. Scale- 0-60. Sweep- 0.00 to 5.00.
2. For general programming- Select an FIO2 of 60%. THEN:
a). Rotate all Alpha control knobs until their arrows are under their 12:00 indexes except Yellow DEMAND CPAP/PEEP control knob rotated full (clockwise).
b). Black INSPIRATORY TIME- control knob Arrow rotated under the 12:00 index for about a 2 second Inspiratory Interval.
c). Black EXPIRATORY TIME- control knob Arrow rotated under the 12:00 index for about a 2 second Expiratory Interval.
d). Green OSCILLATORY CPAP/PEEP- control knob rotated (counterclockwise) until 3:00 is under the 12:00 index position.
e). Rotate Master switch to- ON.
RESET MONITRON® ALARM
OBSERVE: A two step increase in oscillatory (PIP) on the Monitron Screen.
a). First stage is the OSCILLATORY BASELINE created by an Oscillatory Demand CPAP selection of between 5-10 cm H2O.
b). Second stage is created by PERCUSSIVE FLOWRATE selection of about 25-35 cm H2O.
Note: It is recommended NOT to use yellow DEMAND CPAP/PEEP during VDR programming. If elected, do not use more than 4 cm H2O of DEMAND CPAP/PEEP during OSCILLATORY DEMAND CPAP programming. Non-oscillatory, DEMAND CPAP/PEEP is generally reserved for pressure support during spontaneous breathing to reduce the work of breathing.
Note: If DEMAND CPAP is selected with a programmed OSCILLATORY DEMAND CPAP, it will program an accumulative increase.